How Number of Control Points Influences the Dynamic IMRT Plan Quality and Deliverability

no image available
s sharma

s sharma1*, D Manigandan2 , S Chander1 , V Subramani1 , P Julka1 , G Rath1 , (1) All India Institute of Medical Sciences (AIIMS), New Delhi, Delhi, (2) Sri Siddhivinayak Ganapathi Cancer hospital, Miraj, Maharastra,

Presentations

SU-E-T-199 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose:To study the influence of number of control points on plan quality and deliverability.

Methods:Five previously treated patients of carcinoma of rectum were selected. Planning target volume (PTV) and organs at risk (OARs) i.e. bladder and bowel were contoured. Dynamic IMRT plans (6MV, 7-fields, 45Gy/25 fractions and prescribed at 95% isodose) were created in Eclipse (Varian medical system, Palo Alto, CA) treatment planning system (TPS) for Varian CL2300C/D linear-accelerator. Base plan was calculated with 166 control points, variable mode (Eclipse Default). For generating other plans, all parameters were kept constant, only number of control points (Fixed mode) was varied as follows: 100, 166 and 200. Then, plan quality was analyzed in terms of maximum and mean dose received by the PTV and OARs. For plan deliverability, TPS calculated fluence was verified with I’matriXX (IBA Dosimetry, Germany) array and compared with TPS dose-plane using gamma index criteria of 3% dose difference and 3mm distance to agreement (DTA). Total number of monitor units (MU) required to deliver a plan was also noted.

Results:The maximum variation for the PTV maximum with respect to eclipse default control point (166) was 0.28% (0.14Gy). Similarly, PTV mean varied only up to 0.22 %( 0.11Gy). Bladder maximum and bladder mean varied up to 0.51% (0.24Gy) and 0.16% (0.06Gy). The variation for the bowel maximum and bowel mean was also only 0.39% (0.19Gy) and 0.33% (0.04Gy). Total MU was within 0.32 % (4MU). Average gamma pass rate using different control points for five patients are 98.75±0.33%, 99.37±0.09%, 99.29±0.12%, 98.14±0.13% and 99.25±0.14% respectively.

Conclusion:Slight variation (<1%) in PTV and OARs maximum and mean doses was observed with varying number of control points. Monitor unit was also not varied much. Reducing number of control points did not showed any comprise in plan deliverability in terms of gamma index pass rate.