Impact of the Intra and Inter Observer Variability in the Delineation of Parotid Glands On the Dose Calculation During Head and Neck Helical Tomotherapy


A Jodda

A Jodda*, T Piotrowski , Greater Poland Cancer Centre Poznan

Presentations

SU-E-J-275 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose:
The intra- and inter-observer variability in delineation of the parotids on the kilo-voltage computed tomography (kVCT) and mega-voltage computed tomography (MVCT) were examined to establish their impact on the dose calculation during adaptive head and neck helical tomotherapy (HT).

Methods:
Three observers delineated left and right parotids for ten randomly selected patients with oropharynx cancer treated on HT. The pre-treatment kVCT and the MVCT from the first fraction of irradiation were selected to delineation. The delineation procedure was repeated three times by each observer. The parotids were delineated according to the institutional protocol. The analyses included intra-observer reproducibility and inter-structure, -observer and -modality variability of the volume and dose.

Results:
The differences between the left and right parotid outlines were not statistically significant (p>0.3). The reproducibility of the delineation was confirmed for each observer on the kVCT (p>0.2) and on the MVCT (p>0.1). The inter-observer variability of the outlines was significant (p<0.001) as well as the inter-modality variability (p<0.006). The parotids delineated on the MVCT were 10% smaller than on the kVCT. The inter-observer variability of the parotids delineation did not affect the average dose (p=0.096 on the kVCT and p=0.176 on the MVCT). The dose calculated on the MVCT was higher by 3.3% than dose from the kVCT (p=0.009).

Conclusion:
Usage of the institutional protocols for the parotids delineation reduces intra-observer variability and increases reproducibility of the outlines. These protocols do not eliminate delineation differences between the observers, but these differences are not clinically significant and do not affect average doses in the parotids. The volumes of the parotids delineated on the MVCT are smaller than on the kVCT, which affects the differences in the calculated doses.