Quality Assurance for Gamma Knife Perfexion Using the Exradin W1 Plastic Scintillation Detector

no image available
F Therriault-Proulx

R Pino1 , F Therriault-Proulx2*, J Yang2 , S Beddar2 , (1) Houston Methodist Hospital (2) University of Texas MD Anderson Cancer Center, Houston, Texas

Presentations

SU-E-T-476 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose:
To perform dose profile and output factor measurements for the Exradin W1 plastic scintillation detector (PSD) for the Gamma Knife Perfexion (GKP) collimators in a Lucy phantom and to compare these values to an Exradin A16 ion chamber, EBT3 radiochromic film and treatment planning system (TPS) data.

Methods:
We used the Exradin W1 PSD which has a small volume, near-water equivalent sensitive element. It has also been shown to be energy independent. This new detector is manufactured and distributed by Standard Imaging, Inc. Measurements were performed for all three collimators (4 mm, 8 mm and 16 mm) for the GKP. The Lucy phantom with the PSD inserted was moved in small steps to acquire profiles in all three directions. EBT3 film was inserted in the Lucy phantom and exposed to a single shot for each collimator. Relative output factors were measured using the three detectors while profiles acquired with the PSD were compared to the ones measured with EBT3 radiochromic film.

Results:
Measured output factors relative to the largest collimator are as follows
Collimator PS EBT3 A16
16mm 1.000 1.000 1.000
8mm 0.892 0.881 0.883
4mm 0.795 0.793 0.727
The nominal (vendor) OFs for GKP are 1.000, 0.900, and 0.814, for collimators 16 mm, 8 mm and 4 mm, respectively. There is excellent agreement between all profiles measured with the PSD and EBT3 as well as with the TPS data provided by the vendor.

Conclusion:
Output factors measured with the W1 were consistent with the ones measured with EBT3 and A16 ion chamber. Measured profiles are in excellent agreement. The W1 detector seems well suited for beam QA for Gamma Knife due to its dosimetric characteristics.

Funding Support, Disclosures, and Conflict of Interest: Sam Beddar would like to disclose a NIH/NCI SBIR Phase II grant (2R44CA153824-02A1) with Standard Imaging, Title: "Water-Equivalent Plastic Scintillation Detectors for Small Field Radiotherapy"