Evaluation of Four Commercial Dosimetry Systems for Routine Patient-Specific Tomotherapy Delivery Quality Assurance
A Xing*, S Arumugam , S Deshpande , A George , L Holloway , P Vial , G Goozee , Liverpool Cancer Therapy Centre, Sydney, NSW
Presentations
SU-E-T-407 Sunday 3:00PM - 6:00PM Room: Exhibit HallPurpose:The purpose of this project was to evaluate the performance of four commercially available dosimetry systems for Tomotherapy delivery quality assurance (DQA).
Methods:Eight clinical patient plans were chosen to represent a range of treatment sites and typical clinical plans. Four DQA plans for each patient plan were created using the TomoTherapy DQA Station (Hi-Art version 4.2.1) on CT images of the ScandiDose Delta4, IBA MatriXX Evolution , PTW Octavius 4D and Sun Nuclear ArcCHECK phantoms. Each detector was calibrated following the manufacture-provided procedure. No angular response correction was applied. All DQA plans for each detector were delivered on the Tomotherapy Hi-Art unit in a single measurement session but on different days. The measured results were loaded into the vendor supplied software for each QA system for comparison with the TPS-calculated dose. The Gamma index was calculated using 3%/3mm, 2%/2mm with 10% dose threshold of maximum TPS calculated dose.
Results:Four detector systems showed comparable gamma pass rates for 3%/3m, which is recommended by AAPM TG119 and commonly used within the radiotherapy community. The averaged pass rates ± standard deviation for all DQA plans were (98.35±1.97)% for ArcCHECK, (99.9%±0.87)% for Matrix, (98.5%±5.09)% for Octavius 4D, (98.7%±1.27)% for Delata4. The rank of the gamma pass rate for individual plans was consistent between detectors. Using 2%/2mm Gamma criteria for analysis, the Gamma pass rate decreased on average by 9%, 8%, 6.6% and 5% respectively. Profile and Gamma failure map analysis using the software tools from each dosimetry system indicated that decreased passing rate is mainly due to the threading effect of Tomo plan.
Conclusion:Despite the variation in detector type and resolution, phantom geometry and software implementation, the four systems demonstrated similar dosimetric performance, with the rank of the gamma pass rate consistent for the plans considered.