Experimental Determination of K Factor in Small Field Dosimetry
I Das1*, Y Akino1 , P Francescon2 , (1) Indiana University- School of Medicine, Indianapolis, IN, (2) Ospedale Di Vicenza, Vicenza,
Presentations
TU-F-BRE-5 Tuesday 4:30PM - 6:00PM Room: Ballroom EPurpose: Small-field dosimetry is challenging due to charged-particle disequilibrium, source occlusion and more importantly finite size of detectors. IAEA/AAPM has published approach to convert detector readings to dose by k factor. Manufacturers have been trying to provide various types of micro-detectors that could be used in small fields. However k factors depends on detector perturbations and are derived using Monte Carlo simulation. PTW has introduced a microDiamond for small-field dosimetry. An experimental approach is presented to derive the k factor for this detector.
Methods: PTW microDiamond is a small volume detector with 1.1 mm radius and 1.0 micron thick synthetic diamond. Output factors were measured from 1x1cm2 to 12x12 cm2 on a Varian machine at various depths using various micro-detectors with published k factors. Dose is calculated as reading * K. Assuming k factor is accurate, output factor should be identical with every micro-detectors. Hence published k values (Francescon et al Med Phys 35, 504-513,2008) were used to covert readings and then output factors were computed. Based on the converged curve from other detectors, k factor for microDiamond was computed versus field size.
Results:Traditional output factors as ratio of readings normalized to 10x10 cm2 differ significantly for micro-detectors for fields smaller than 3x3 cm2 which are now being used extensively. When readings are converted to dose, the output factor is independent of detector. Based on this method, k factor for microDiamond was estimated to be nearly constant 0.993±0.007 over varied field sizes.
Conclusion: Our method provides a unique opportunity to determine the k factor for any unknown detector. It is shown that even though k factor depends on machine type due to focal spot, however for fields ≥1x1 cm2 this method provides accurate evaluation of k factor. Additionally microDiamond could be used with assumption that k factor is nearly unity.