Commissioning and Comprehensive Evaluation of the ArcCHECK Cylindrical Diode Array for VMAT QA
V Chaswal*, M Weldon, N Gupta, Y Rong, Ohio State University, Columbus, OH
Presentations
SU-E-T-631 Sunday 3:00PM - 6:00PM Room: Exhibit HallPurpose:Commissioning and comprehensive evaluation of ArcCHECK phantom for dosimetry of VMAT QA, using 6MV photon beam with and without the flattening filter.
Methods:ArcCHECK was evaluated for response dependency on linac dose rate, instantaneous dose rate, radiation field size, beam angle and couch insertion. Scatter dose characterization, consistency and symmetry of response, dosimetric accuracy of fixed aperture arcs and clinical VMAT plans were investigated. Measurements were done using TrueBeamâ„¢ STx accelerator (Console version 1.6) with a 6 MV beam with and without flattening filter. Reference dose-grids were calculated using Eclipse TPS Analytical Anisotropic Algorithm (AAA version 10.0.39). Planned doses were calculated using symmetric 2mm 3D dose grids with 4 degree angular resolution defaulted to each control point. Gamma evaluations were performed in absolute dose mode, with default normalization to maximum dose in the curved plane and a low dose threshold of 10% to restrict the analysis to clinically relevant areas. Global and local gamma indices at 3mm/3% and 2mm/2% level were computed using SNC software (version 6.0).
Results:Results of gamma analysis demonstrated an overall agreement between ArcCHECK measured and TPS calculated reference doses. Field size dependency was within 0.5% of the reference. Dose-rate based dependency was well within 1% of the TPS reference and the angular dependency was ±3% of the reference, as tested for BEV angles. At the level of 3%/3mm, narrow and wide open arcs as well as clinical VMAT cases demonstrated high level of dosimetry accuracy in global gamma passing rates for both 6X and 6F beams. At the level of 2%/2mm two VMAT cases involving the narrow heavily modulated arcs demonstrated lower passing rates.
Conclusion:ArcCHECK phantom with latest software and hardware upgrades is suitable for VMAT QA. For higher sensitivity of 2%/2mm gamma analysis, we intend to use it as one of the VMAT QA evaluation metrics.
Contact Email: