Encrypted login | home

Program Information

Commissioning and Acceptance Testing of a Commercial Monte Carlo Electron Dose Calculation Model (eMC) for TrueBeam


R Sheu

R Sheu*, T Tseng , A Powers , Y Lo , Mount Sinai Medical Center, New York, NY

Presentations

SU-E-T-71 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: To provide commissioning and acceptance test data of the Varian Eclipse electron Monte Carlo model (eMC v.11) for TrueBeam linac. We also investigated the uncertainties in beam model parameters and dose calculation results for different geometric configurations.

Methods: For beam commissioning, PTW CC13 thimble chamber and IBA Blue Phantom2 were used to collect PDD and dose profiles in air. Cone factors were measured with a parallel plate chamber (PTW N23342) in solid water. GafChromic EBT3 films were used for dose calculation verifications to compare with parallel plate chamber results in the following test geometries: oblique incident, extended distance, small cutouts, elongated cutouts, irregular surface, and heterogeneous layers.

Results: Four electron energies (6e, 9e, 12e, and 15e) and five cones (6x6, 10x10, 15x15, 20x20, and 25x25) with standard cutouts were calculated for different grid sizes (1, 1.5,2, and 2.5 mm) and compared with chamber measurements. The results showed calculations performed with a coarse grid size underestimated the absolute dose. The underestimation decreased as energy increased. For 6e, the underestimation (max 3.3 %) was greater than the statistical uncertainty level (3%) and was systematically observed for all cone sizes. By using a 1mm grid size, all the calculation results agreed with measurements within 5% for all test configurations. The calculations took 21s and 46s for 6e and 15e (2.5mm grid size) respectively distributed on 4 calculation servants.

Conclusion: In general, commissioning the eMC dose calculation model on TrueBeam is straightforward and thedose calculation is in good agreement with measurements for all test cases. Monte Carlo dose calculation provides more accurate results which improves treatment planning quality. However, the normal acceptable grid size (2.5mm) would cause systematic underestimation in absolute dose calculation for lower energies, such as 6e. Users need to be cautious in this situation.


Contact Email: