Geometric and Dosimetric Evaluation of a 4D-CBCT Reconstruction Technique Using Prior Knowledge


Y Zhang

Y Zhang*, F Yin , L Ren , Duke University Medical Center, Durham, NC

Presentations

SU-D-17A-1 Sunday 2:05PM - 3:00PM Room: 17A

Purpose:
To evaluate a 4D-CBCT reconstruction technique both geometrically and dosimetrically

Method:
A prior-knowledge guided 4D-CBCT reconstruction method named the motion-modeling and free-form deformation (MM-FD) has been developed. MM-FD views each phase of the 4D-CBCT as a deformation of a prior CT volume. The deformation field is first solved by principal component analysis based motion modeling, followed by constrained free-form deformation.

The 4D digital extended-cardiac-torso (XCAT) phantom was used for comprehensive evaluation. Based on a simulated 4D planning CT of a lung patient, 8 different scenarios were simulated to cover the typical on-board anatomical and respiratory variations: (1) synchronized and (2) unsynchronized motion amplitude change for body and tumor; tumor (3) shrinkage and (4) expansion; tumor average position shift in (5) superior-inferior (SI) direction, (6) anterior-posterior (AP) direction and (7) SI, AP and lateral directions altogether; and (8) tumor phase shift relative to the respiratory cycle of the body. Orthogonal-view 30° projections were simulated based on the eight patient scenarios to reconstruct on-board 4D-CBCTs.

For geometric evaluation, the volume-percentage-difference (VPD) was calculated to assess the volumetric differences between the reconstructed and the ground-truth tumor.

For dosimetric evaluation, a gated treatment plan was designed for the prior 4D-CT. The dose distributions were calculated on the reconstructed 4D-CBCTs and the ground-truth images for comparison.

The MM-FD technique was compared with MM-only and FD-only techniques.

Results:
The average (±s.d.) VPD values of reconstructed tumors for MM-only, FD-only and MM-FD methods were 59.16%(± 26.66%), 75.98%(± 27.21%) and 5.22%(± 2.12%), respectively. The average min/max/mean dose (normalized to prescription) of the reconstructed tumors by MM-only, FD-only, MM-FD methods and ground-truth tumors were 78.0%/122.2%/108.2%, 13%/117.7%/86%, 58.1%/120.8%/103.6% and 57.6%/118.6%/101.8%,respectively.

Conclusion:
The MM-FD method provides superior reconstruction accuracy both geometrically and dosimetrically, which can potentially be used for 4D target localization, dose tracking and adaptive radiation therapy.