Comparison of Radiation Dose for Radiography and EOS in Adolescent Scoliosis Patients


B Schueler

B Schueler*, A Walz-Flannigan , Mayo Clinic, Rochester, MN

Presentations

SU-E-I-15 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: To estimate patient radiation dose for whole spine imaging using EOS, a new biplanar slot-scanning radiographic system and compare with standard scoliosis radiography.
Methods: The EOS imaging system (EOS Imaging, Paris, France) consists of two orthogonal x-ray fan beams which simultaneously acquire frontal and lateral projection images of a standing patient. The patient entrance skin air kerma was measured for each projection image using manufacturer-recommended exposure parameters for spine imaging. Organ and effective doses were estimated using a commercially-available Monte Carlo simulation program (PCXMC, STUK, Radiation and Nuclear Safety Authority, Helsinki, Finland) for a 15 year old mathematical phantom model. These results were compared to organ and effective dose estimated for scoliosis radiography using computed radiography (CR) with standard exposure parameters obtained from a survey of pediatric radiographic projections.
Results: The entrance skin air kerma for EOS was found to be 0.18 mGy and 0.33 mGy for posterior-anterior (PA) and lateral projections, respectively. This compares to 0.76 mGy and 1.4 mGy for CR, PA and lateral projections. Effective dose for EOS (PA and lateral projections combined) is 0.19 mSv compared to 0.51 mSv for CR.
Conclusion: The EOS slot-scanning radiographic system allows for reduced patient radiation dose in scoliosis patients as compared to standard CR radiography.