Comparison of 6MV Photon Dose in a Perpendicular and Parallel Magnetic Field
A Ghila*, B Fallone , S Rathee , Cross Cancer Institute, Edmonton, AB
Presentations
SU-E-J-38 Sunday 3:00PM - 6:00PM Room: Exhibit HallPurpose: Integrating a linac with an MRI system would allow for real time tumour tracking however the patient will be irradiated in the presence of a magnetic field. The present study experimentally investigates the magnetic field effects on entrance, exit, and interface dose for both transverse and parallel magnetic fields.
Methods: Polystyrene was used to construct a set of phantoms for Gafchromic film measurements. One phantom had an adjustable air gap and four other phantoms had one surface at various angles. The linac-MR prototype consisting of a biplanar permanent magnet coupled to a linac was used for the transverse magnetic field measurements. A couple of solenoid electromagnets, stacked on top of each other and irradiated along their bore, were used for the parallel field measurements.
Results: All doses are relative to no magnetic field. The transverse magnetic field reduced the entrance dose for all surface angles by strongly deflecting the contaminant electrons. The exit dose in a transverse magnetic field was found to be significantly higher. The entrance dose with a parallel magnetic field present is higher due to the contaminant electrons being concentrated within the beam area. The air gap phantom measurements, done in a transverse magnetic field, show a significant increase of the dose at the proximal side of the air gap and a decrease at the distal side. The measurements, done in the parallel magnetic field, show the concentration of secondary electrons in the air gap.
Conclusion: The radiation dose measurements of a 6MV beam in a parallel and transverse magnetic field presented here are currently being replicated using Monte Carlo simulations. This verified Monte Carlo system could provide the dose calculation basis for future linac-MR systems.
Contact Email: