Point Dose Variation Using a Single Ir-192 HDR Brachytherapy Plan for Two Treatments with a Single Tandem-Ovoid Insertion for Cervical Cancer

no image available
X Liang

X Liang*, S Morrill , M Hardee , E Han , J Penagaricano , X Zhang , R Vaneerat , University of Arkansas Medical Science, Little Rock, AR

Presentations

SU-E-T-263 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose:
To evaluate the point dose variations between Ir-192 HDR treatments on two consecutive days using a single tandem-ovoid insertion without replanning in cervical cancer patients.

Methods:
This study includes eleven cervical cancer patients undergoing HDR brachytherapy with a prescribed dose of 28 Gy in 4 fractions. Each patient had two tandem-ovoid insertions one week apart. Each insertion was treated on consecutive days with rescanning and replanning prior to each treatment. To study the effect of no replanning for day 2 treatments, the day 1 plan dwell position and dwell time with decay were applied to the day 2 CT dataset. The point dose variations on the prescription point H (defined according to American Brachytherapy Society), and normal tissue doses at point B, bladder, rectum and vaginal mucosa (based on ICRU Report 38) were obtained.

Results:
Without replanning, the mean point H dose variation was 4.6 ± 10.7% on the left; 2.3 ± 2.9% on the right. The mean B point variation was 3.8 ± 4.9% on the left; 3.6 ± 4.7% on the right. The variation in the left vaginal mucosal point was 12.2 ± 10.7%; 9.5 ± 12.5% on the right; the bladder point 5.5 ± 7.4%; and the rectal point 7.9 ± 9.1%.

Conclusion:
Without replanning, there are variations both in the prescription point and the normal tissue point doses. The latter can vary as much as 10% or more. This is likely due to the steep dose gradient from brachytherapy compounded by shifts in the positions of the applicator in relationship to the patients anatomy. Imaging prior to each treatment and replanning ensure effective and safe brachytherapy are recommended.