Program Information
Dose Verification for Accuboost Applicators Using TLD, Ion Chamber and Gafchromic Film Measurements
W Chisela*, R Yao , G Dorbu , Columbus Regional Healthcare, Columbus, GA
Presentations
SU-E-T-118 Sunday 3:00PM - 6:00PM Room: Exhibit HallPurpose: To verify dose delivered with HDR Accuboost applicators using TLD, ion chamber and Gafchromic film measurements and to examine applicator leakage.
Methods: A microSelectron HDR unit was used to deliver a dose of 50cGy to the mid-plane of a 62mm thick solid water phantom using dwell times from Monte Carlo pre-calculated nomograms for a 60mm, 70mm Round and 60mm Skin-Dose Optimized (SDO) applicators respectively. GafChromic EBT3+ film was embedded in the phantom mid-plane horizontally to measure dose distribution. Absolute dose was also measured with TLDs and an ADCL calibrated parallel-plate ion chamber placed in the film plane at field center for each applicator. The film was calibrated using 6MV x-ray beam. TLDs were calibrated in a Cs-137 source at UW-Madison calibration laboratory. Radiation leakage through the tungsten alloy shell was measured with a film wrapped around outside surface of a 60mm Round applicator.
Results: Measured maximum doses at field center are consistently lower than predicated by 5.8% for TLD, 8.8% for ion chamber, and 2.6% for EBT3+ film on average, with measurement uncertainties of 2.2%, 0.3%, and 2.9% for TLD, chamber, film respectively. The total standard uncertainties for ion chamber and Gafchromic film measurement are 4.9% and 4.6% respectively[1]. The area defined by the applicator aperture was covered by 80% of maximum dose for 62mm compression thickness. When 100cGy is delivered to mid-plane with a 60mm Round applicator, surface dose ranges from 60cGy to a maximum of 145cGy, which occurs at source entrance to the applicator.
Conclusion: Measured doses by all three techniques are consistently lower than predicted in our measurements. For a compression thickness of 62 mm, the field size defined by the applicator is only covered by 80% of prescribed dose. Radiation leakage of up to 145cGy was found at the source entrance of applicators.
Contact Email: