Development of An Abdominal Compression Device for Respiratory Correlated Radiation Therapy

no image available
T Kim

T Kim1*, S Kang1 , D Kim1 , S Kim2 , T Suh1 , (1) Catholic Univ Medical College, Seoul, Seoul, (2) Virginia Commonwealth University, Richmond, Virginia

Presentations

SU-E-J-48 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose:The aim of this study is to develop the abdominal compression device which could control pressure level according to the abdominal respiratory motion and evaluate its feasibility.

Methods:In this study, we focused on developing the abdominal compression device which could control pressure level at any point of time so the developed device is possible to use a variety of purpose (gating technique or respiratory training system) while maintaining the merit of the existing commercial device. The compression device (air pad form) was designed to be able to compress the front and side of abdomen and the pressure level of the abdomen is controlled by air flow. Pressure level of abdomen (air flow) was determined using correlation data between external abdominal motion and respiratory volume signal measured by spirometer. In order to verify the feasibility of the device, it was necessary to confirm the correlation between the abdominal respiratory motion and respiratory volume signal and cooperation with respiratory training system also checked.

Results:In the previous study, we could find that the correlation coefficient ratio between diaphragm and respiratory volume signal measured by spirometer was 0.95. In this study, we confirmed the correlation between the respiratory volume signal and the external abdominal motion measured by belt-transducer (correlation coefficient ratio was 0.92) and used the correlated respiratory volume data as an abdominal pressure level. It was possible to control the pressure level with negligible time delay and respiratory volume data based guiding waveforms could be properly inserted into the respiratory training system.

Conclusion:Through this feasibility study, we confirmed the correlation between the respiratory volume signal and the external abdominal motion. Also initial assessment of the device and its compatibility with the respiratory training system were verified. Further study on application in respiratory gated therapy and respiratory training system will be investigated.

Funding Support, Disclosures, and Conflict of Interest: This work was supported by Radiation Technology R&D program (No. 2013M2A2A7043498)and Basic Atomic Energy Research Institute (BAERI)(No. NRF-2009-0078390) through the National Research Foundation of Korea funded by the Ministry of Science, ICT&Future Planning.