Planning and Delivery of the Fully Dynamic Trajectory Modulated Arc Therapy: Application to Accelerated Partial Breast Irradiation


J Liang

J Liang1*, T Atwood1, B Fahimian1, E Chin1, K Otto2, D Hristov1, (1)Department of Radiation Oncology, Stanford University, CA, (2)Department of Physics, University of British Columbia, BC

Presentations

TH-C-12A-9 Thursday 10:15AM - 12:15PM Room: 12A

Purpose: A novel trajectory modulated arc therapy (TMAT) system was developed that uses source motion trajectory involving synchronized gantry rotation with translational and rotational couch movement. MLC motion and dose rate were fully optimized for dynamic beam delivery. This work presents a platform for planning deliverable TMAT on a collision free coronal trajectory and evaluates its benefit for accelerated partial breast irradiation (APBI) in a prone position.

Methods: The TMAT algorithm was built on VMAT with modifications (physical properties on couch movement were defined) and enhancements (pencil beam dose calculation engine to support extended SSDs) to make it feasible for TMAT delivery. A Matlab software environment for TMAT optimization and dose calculation was created to allow any user specified motion axis. TMAT delivery was implemented on Varian TrueBeamTM STx via XML scripts. 10 prone breast irradiation cases were evaluated in VMAT and compared with a 6-field non-coplanar IMRT plan. Patient selection/exclusion criteria and structure contouring followed the guidelines of NSABP B-39/RTOG 0413 protocol.

Results: TMAT delivery time was ~4.5 minutes. 251.5°±7.88° of non-isocentric couch arc was achieved by the optimized trajectory with 180-210 control points at 1°-2° couch increments. The improved dose distribution by TMAT was most clearly observed by the marked reduction in the volume of irradiated normal breast tissue in the high dose region. The ratios of the normal breast tissue volume receiving more than 50%, 80% and 100% of the prescription dose for TMAT versus IMRT were: V50%(TMAT/IMRT) = 78.38%±13.03%, V80%(TMAT/IMRT) = 44.19%±9.04% and V100% (TMAT/IMRT) = 9.96%±7.55%, all p≤0.01.

Conclusion: The study is the first demonstration of planning and delivery implementation of a fully dynamic APBI TMAT system with continuous couch motion. TMAT achieved significantly improved dosimetry over non-coplanar IMRT on dose volume parameters correlated with toxicity and cosmetic outcome of APBI.


Funding Support, Disclosures, and Conflict of Interest: This project was supported by Varian Research Grant.