Size-Specific Dose Estimate (SSDE) Provides a Simple Method to Calculate Organ Dose for Pediatric CT Examinations


S Brady

B Moore1 , S Brady1*, A Mirro2 , R Kaufman1 , (1) St Jude Children's Research Hospital, Memphis, TN, (2) Washington University, St. Louis, MO

Presentations

MO-E-17A-4 Monday 1:45PM - 3:45PM Room: 17A

Purpose: Investigate the correlation of SSDE with organ dose in a pediatric population.

Methods: Four anthropomorphic phantoms, representing a range of pediatric body habitus, were scanned with MOSFET dosimeters placed at 23 organ locations to determine absolute organ dosimetry. Phantom organ dosimetry was divided by phantom SSDE to determine correlation between organ dose and SSDE. Correlation factors were then multiplied by patient SSDE to estimate patient organ dose. Patient demographics consisted of 352 chest and 241 abdominopelvic CT examinations, 22 ± 15 kg (range 5-55 kg) mean weight, and 6 ± 5 years (range 4 mon to 23 years) mean age. Patient organ dose estimates were compared to published pediatric Monte Carlo study results.

Results: Phantom effective diameters were matched with patient population effective diameters to within 4 cm. 23 organ correlation factors were determined in the chest and abdominopelvic region across nine pediatric weight subcategories. For organs fully covered by the scan volume, correlation in the chest (average 1.1; range 0.7-1.4) and abdominopelvic (average 0.9; range 0.7-1.3) was near unity. For organs that extended beyond the scan volume (i.e., skin, bone marrow, and bone surface), correlation was determined to be poor (average 0.3; range: 0.1-0.4) for both the chest and abdominopelvic regions, respectively. Pediatric organ dosimetry was compared to published values and was found to agree in the chest to better than an average of 5% (27.6/26.2) and in the abdominopelvic region to better than 2% (73.4/75.0).

Conclusion: Average correlation of SSDE and organ dosimetry was found to be better than ± 10% for fully covered organs within the scan volume. This study provides a list of organ dose correlation factors for the chest and abdominopelvic regions, and describes a simple methodology to estimate individual pediatric patient organ dose based on patient SSDE.