Design and Development of Isocenter Parameter System for CT Simulation Laser Based On DICOM RT

no image available
G Luo

G Luo1*, (1)Sun Yat-sen University Cancer Center; State Key Laboratory of Oncology in South China

Presentations

SU-E-T-99 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: In order to receive DICOM files from treatment planning system and generate patient isocenter positioning parameter file for CT laser system automatically, this paper presents a method for communication with treatment planning system and calculation of isocenter parameter for each radiation field.

Methods:Coordinate transformation and laser positioning file formats were analyzed, isocenter parameter was calculated via data from DICOM CT Data and DICOM RTPLAN file. An in-house software-DicomGenie was developed based on the object-oriented program platform-Qt with DCMTK SDK (Germany OFFIS company DICOM SDK) . DicomGenie was tested for accuracy using Philips CT simulation plan system (Tumor LOC, Philips) and A2J CT positioning laser system (Thorigny Sur Marne, France).

Results:DicomGenie successfully established DICOM communication between treatment planning system, DICOM files were received by DicomGenie and patient laser isocenter information was generated accurately. Patient laser parameter data files can be used for for CT laser system directly.

Conclusion:In-house software DicomGenie received and extracted DICOM data, isocenter laser positioning data files were created by DicomGenie and can be use for A2J laser positioning system.