Feasibility of Using a Grid to Detect and Correct the Geometric Variations in Flat-Panel Based Cone Beam CT
G Zou1*, L Ren2, J Kim1, D Jaffray3, I Chetty1, J Jin1, (1) HENRY FORD HEALTH SYSTEM, Detroit, MI, (2) Duke University Medical Center, Durham, NC, (3) Princess Margaret Cancer Centre, Toronto, ON
WE-G-141-7 Wednesday 4:30PM - 6:00PM Room: 141Purpose: To study the feasibility of using a grid, shown previously to mitigate the scatter problem, to detect and correct geometric variations in flat-panel-based cone beam CT (CBCT) imaging.
Methods: The study was performed on the CBCT system of a Varian Trilogy linac. The grid was located before the imaging object and attached to the bowtie-filter with a source-to-grid distance (SGD) of 40-50 cm. We first studied the correlations in geometric variation between the grid, bowtie-filter, source and imager. The grid position was represented by BBs in a plastic-plate replacing the grid. The bowtie-filter position was determined by a BB placed on its surface. The source and imager geometries were derived from the projections of 16 BBs in an IsoCal phantom using an in-house-developed computer program. Multiple scans were performed with different SGD in a 2-month period. Correlations were analyzed and used to predict the geometric variations of the bowtie-filter, source and imager from the grid variation obtained using a grid in a CatPhan scan.
Results: The grid and bowtie-filter showed exactly same variations (up to 5 mm) in Y-direction during a 360° rotation, with a sine-like pattern superimposed by 1-2 mm random vibrations. The variation patterns were also sine-like curves in the X-direction. However, they were smooth and repeatable, with differences in phase and amplitude between the two, and the amplitude varying with SGD. The source and imager showed similar and repeatable 0.3-0.5 mm sine-like variation patterns. Based on these results, a model was developed to predict and correct the geometric variations in a CBCT scan with the grid.
Conclusion: The preliminary results suggest that it is able to predict geometric variations of the CBCT system. Future study is underway to verify whether correction of these variations will improve the spatial resolution and mitigate the bowtie-filter variation-induced crescent artifact.
Funding Support, Disclosures, and Conflict of Interest: The study is supported by NIH Grant Number 5 R01 CA166948-02.