Monte Carlo Determination of WAFAC Corrections for the Canadian LDR Primary Standard
E Mainegra*, H Shen, J McCaffrey, National Research Council, Ottawa, ON
SU-E-T-32 Sunday 3:00PM - 6:00PM Room: Exhibit HallPurpose: Determine correction factors for a wide angle free-air chamber (WAFAC) using Monte Carlo techniques for the realization of the air-kerma strength, Sk, for low dose rate (LDR) ¹²⁵I and ¹⁰³Pd brachytherapy sources. This WAFAC will become the Canadian primary standard for LDR brachytherapy sources.
Methods: The EGSnrc user-code egs_fac is used for estimating the correction factors required to determine the air-kerma at the WAFACs point of measurement (POM) using MC simulation for a ¹²⁵I and a ¹⁰³Pd source. The geometry library egs++ allows simulation of the sources and the WAFAC in detail. Spectra are taken from the most recently published data. Air and aluminum attenuation curves for the ¹²⁵I seed were also measured with the WAFAC.
Results: Correction values compared with those reported by NIST for its LDR primary standard (of very similar design) are in excellent agreement when using the NIST approach of assuming an isotropic point source and using experimental main spectrum lines. Differences in both methodologies for determining the WAFAC corrections are discussed and their impact evaluated. The effect of the source and the holder on the scatter correction is quantified separately. The calculations resulted in a modification of the design to reduce collimator transmission. A 3.2 mm lead sheet was added behind the WAFAC aperture, attenuating the ¹²⁵I and the ¹⁰³Pd photon beams to less than 0.02 %. MC attenuation curves for air and aluminum for the ¹²⁵I seed are in good agreement with experimental data. The largest correction is the aluminum filter attenuation which can be as large as 4 %.
Conclusion: Despite an inconsistency in the scatter estimation, the NIST approximation has little impact on the determination of Sk. The scatter correction is sensitive to the collimation system dimensions. Using MC-based air and aluminum attenuation corrections for the ¹²⁵I seed is experimentally justified.
Contact Email: