A Prompt Gamma-Ray Spectroscopy System for Clinical Studies of in Vivo Proton Range Verification


J Verburg

J Verburg*, T Bortfeld , Massachusetts General Hospital and Harvard Medical School, Boston, MA

Presentations

TU-FG-BRB-6 (Tuesday, August 2, 2016) 1:45 PM - 3:45 PM Room: Ballroom B


Purpose: We present a new system to perform prompt gamma-ray spectroscopy during proton pencil-beam scanning treatments, which enables in vivo verification of the proton range. This system will be used for the first clinical studies of this technology.

Methods: After successful pre-clinical testing of prompt gamma-ray spectroscopy, a full scale system for clinical studies is now being assembled. Prompt gamma-rays will be detected during patient treatment using an array of 8 detector modules arranged behind a tungsten collimator. Each detector module consists of a lanthanum(III) bromide scintillator, a photomultiplier tube, and custom electronics for stable high voltage supply and signal amplification. A new real-time data acquisition and control system samples the signals from the detectors with analog-to-digital converters, analyses events of interest, and communicates with the beam delivery systems. The timing of the detected events was synchronized to the cyclotron radiofrequency and the pencil-beam delivery. Range verification is performed by matching measured energy- and time-resolved gamma-ray spectra to nuclear reaction models based on the clinical treatment plan. Experiments in phantoms were performed using clinical beams in order to assess the performance of the systems.

Results: The experiments showed reliable real-time analysis of more than 10 million detector events per second. The individual detector modules acquired accurate energy- and time-resolved gamma-ray measurements at a rate of 1 million events per second, which is typical for beams delivered with a clinical dose rate. The data acquisition system successfully tracked the delivery of the scanned pencil-beams to determine the location of range deviations within the treatment field.

Conclusion: A clinical system for proton range verification using prompt gamma-ray spectroscopy has been designed and is being prepared for use during patient treatments. We anticipate to start a first clinical study in the near future.